ZNB-XAI I
Specifications
Model ZNB- XAⅡ
Pumping Mechanism Curvilinear per i st al t i c
I V Set Compatible with IV set s of any standard
Flow Rate 0. 1- 1500 ml / h (in 0. 1 ml/h increments)
Purge, Bolus 100-1500 ml/h (in 0.1 ml/h increments) Purge when pump
stops, bolus when pump starts
Accuracy ±3%
VTBI 1- 20000 ml (in 0. 1 ml increments)
Infusion Mode ml/h, drop/min, time- based, body weight, nutrition
KVO Rate 0.1- 5 ml/h (in 0.1ml/h increments)
Alarms Occlusion, air-in-line, door open, end program, low battery,
end battery, AC power off, motor malfunction, system
malfunction, reminder alarm
Additional Features Real-time infused volume, automatic power switching,
mute button, purge, bolus, system memory, key locker,
change flow rate without stopping the pump
Drug Library Available
Occlusion Sensitivity High, medium, low
History Log 50000 events
Air – in- line Detection Ultrasonic detector
Wireless Management Optional
Drop Sensor Optional
Vehicle Power Available
Power Supply AC: 110/230 V (optional), 50/60 Hz, 20VA Battery:
9.6±1.6V, rechargeable Vehicle power: 12±1.2 V
Battery Life 5 hours at 25 ml/h
Working Temperature 10- 30℃
Relative Humidity 30- 75%
Atmospheric Pressure 860- 1060 hpa
Size 130* 145* 228 mm
Weight 2. 5 kg
Safety Classification Class Ⅰ, type CF